Vacuum Fine-Needle Aspiration Biopsy: Efficiency of the Method from the Perspective of Cytological Assessment — an Experimental Study on Porcine Liver
https://doi.org/10.37174/2587-7593-2025-8-4-86-95
Abstract
Introduction: Morphological veriĮcation remains the cornerstone of tumor diagnosis, providing an objective basis for treatment planning. Among instrumental diagnostic methods, Įne-needle aspiration biopsy (FNA) occupies a leading position due to its minimal invasiveness and high diagnostic yield. However, traditional aspiration techniques are limited by unstable negative pressure and frequent cellular damage, which may compromise smear quality. The development of vacuum Įne-needle aspiration biopsy (v-FNA) introduces the potential for procedural standardization and improved cytological specimen quality.
Objective: To experimentally evaluate the eĸciency and diagnostic value of v-FNA for obtaining cytological material from porcine liver tissue, to assess the feasibility of using small-diameter needles (including 25G), and to compare the results with conventional FNA.
Materials and Methods: The study was conducted on porcine liver specimens, which closely resemble human liver in structure and density. Biopsies were performed using Chiba needles of 18G, 20G, 22G, and 25G diameters. Each puncture was carried out both with traditional syringe aspiration and with vacuum aspiration under a stable negative pressure of -0.8 bar. All procedures were performed under ultrasound guidance to ensure precise needle positioning and controlled advancement through the parenchyma. Cytological smears were prepared using the standard technique and stained by the May–Grunwald method for microscopic evaluation.
Results: The quality and quantity of aspirated material strongly depended on the needle diameter. Conventional FNA using 18G and 20G needles yielded excessive cell masses, often accompanied by hemorrhagic contamination and cellular destruction. Needles of 22G produced more homogeneous and diagnostically suitable smears. The application of vacuum-assisted aspiration signiĮcantly improved cell preservation and smear uniformity. Notably, even 25G needles, traditionally considered ineffective for cytological sampling, provided limited but diagnostically valuable material when used with v-FNA.
Conclusions: Vacuum-assisted fine-needle aspiration biopsy represents a promising direction in cytological diagnostics, improving the quality and informativeness of aspirated material while maintaining minimal invasiveness. The method expands the range of usable needle diameters, supports procedural standardization, and has strong potential for clinical implementation in liver, pancreatic, and lymph node lesion diagnostics.
Keywords
About the Authors
A. O. TagilRussian Federation
Anton O. Tagil
+7-910-786-52-69
40 Frunze str, Smolensk, 214006
Competing Interests:
Not declared
A. V. Borsukov
Russian Federation
Aleksey V. Borsukov
40 Frunze str, Smolensk, 214006
Competing Interests:
Not declared
K. S. Maksimova
Russian Federation
Kseniya S. Maksimova
40 Frunze str, Smolensk, 214006
Competing Interests:
Not declared
V. A. Vorobyova
Russian Federation
Viktoria A. Vorobyova
40 Frunze str, Smolensk, 214006
Competing Interests:
Not declared
References
1. Al-Abbadi MA. Basics of cytology. Avicenna J Med. 2011;1(1):18-28. https://doi.org/10.4103/2231-0770.83719. PMID: 23923127.
2. Dhiman B, Kundu R, Mitra S et al. Hepatocellular carcinoma: Morphological spectrum and subtyping as per the World Health Organization classiĮcation on FNA biopsy with cell block samples. Cancer Cytopathol. 2025;133(4):e70009. https://doi.org/10.1002/cncy.70009. PMID: 40045688.
3. Geijer M, Domanski HA. Targeted Įne-needle aspiration biopsy. In: Fine-Needle Aspiration Biopsy. Cytological Atlas. 2024. p.59. (In Russ.)
4. Tian F, Liu D, Wei N, et al. Prediction of tumor origin in cancers of unknown primary origin with cytology-based deep learn ing. Nat Med. 2024;30(6):1309-19. https://doi.org/10.1038/s41591-024-02915-w. PMID: 38945312.
5. Suzuki R, Shin D, Richards-Kortum R, Coghlan L, Bhutani MS. In vivo cytological observation of liver and spleen using a high-resolution microendoscopy system under endoscopic ultrasound guidance: A preliminary study using a swine model. Endosc Ultrasound. 2016;5(4):239-42. https://doi.org/10.4103/23039027.187867. PMID: 27503155; PMCID: PMC4989404.
6. Busko EA, Mortada VV, Krivorotko PV et al. Breast neoplasms with uncertain malignant potential (B3): Experience of vacuum-assisted biopsy under ultrasound guidance. Radiation Diagnostics and Therapy. 2022;13(3):43-50. (In Russ.) https://doi.org/10.22328/2079-5343-2022-13-3-43-50.
7. Levanov AV, Marushchak EA, Sidorov AO et al. Vacuum aspiration biopsy: Evolution of the method and prevention of hemorrhagic complications. Ambulatory Surgery. 2024;21(2):142-52. (In Russ.) https://doi.org/10.21518/akh2024-021
8. Maksimov KV, Petrova NA, Ivanova LV, et al. Comparison of intermediate results of multifocal and vacuum-assisted biopsy of the breast in evaluating the pathomorphological response after neoadjuvant chemotherapy. Problems in Oncology. 2023;69(6):1057-64. (In Russ.)
9. Ministry of Health of the Russian Federation. Hepatocellular carcinoma: Clinical guidelines. 2025. Available from: https://cr.minzdrav.gov.ru/view-cr/1_4. (In Russ.)
10. Smetanina SV. SpeciĮc features of differential cytological diagnosis of primary and metastatic liver carcinomas. Clinical Laboratory Diagnostics. 2021;66(6):364-70. (In Russ.)
11. Tagil AO, Borsukov AV. Device for performing vacuum Įneneedle aspiration biopsy under ultrasound guidance. Patent RF №2757525 C1. 2021. (In Russ.)
12. Tagil AO, Borsukov AV. Method for increasing the amount of tissue and cytological material during vacuum aspiration biopsy. Patent RF №2770783 C1. 2022. (In Russ.)
Review
For citations:
Tagil A.O., Borsukov A.V., Maksimova K.S., Vorobyova V.A. Vacuum Fine-Needle Aspiration Biopsy: Efficiency of the Method from the Perspective of Cytological Assessment — an Experimental Study on Porcine Liver. Journal of oncology: diagnostic radiology and radiotherapy. 2025;8(4):86-95. (In Russ.) https://doi.org/10.37174/2587-7593-2025-8-4-86-95
JATS XML